Sale!

AX460/100010/STD Single and dual input analyzers for pH/Redox

AX400 analyzers incorporate the latest technology to provide highly reliable, flexible, feature-packed devices that satisfy a diverse range of process monitoring and control applications.
The complete range encompasses solutions for pH/Redox (ORP), conductivity and dissolved oxygen.
AX46x analyzers enable continuous measurements of one or two pH or Redox (ORP) points with simultaneous local display and retransmission. AX46x is used with glass, enamel and
antimony pH electrodes and metal Redox sensors, providing measurements with exceptional accuracy and performance.
AX400 series analyzers are available for either wall-/pipe- or panel-mounting and are rated to IP65.
Categories: ,

Description

Overview


Essential details:AX460/100010/STD Single and dual input analyzers for pH/Redox

AX400 analyzers incorporate the latest technology to provide highly reliable, flexible, feature-packed devices that satisfy a diverse range of process monitoring and control applications.
The complete range encompasses solutions for pH/Redox (ORP), conductivity and dissolved oxygen.
AX46x analyzers enable continuous measurements of one or two pH or Redox (ORP) points with simultaneous local display and retransmission. AX46x is used with glass, enamel and
antimony pH electrodes and metal Redox sensors, providing measurements with exceptional accuracy and performance.
AX400 series analyzers are available for either wall-/pipe- or panel-mounting and are rated to IP65.

An advanced function card provides an additional two current outputs and two further alarm relays that can be assigned to either measured values or sample temperature.
ABB plug-and-produce software automatically reconfigures the analyzer if an option board is added later. No user programming is necessary.
A real-time clock and logbook are also included, making the full-facility versions extremely powerful and versatile.

All versions are supplied with two, fully isolated current outputs as standard, that can be assigned to the measured parameter, sample temperature or any appropriate calculated variables.
Three programmable relay set points are available that can also be assigned as required.
Innovative features such as power-saving display and a diagnostic current output option all contribute to a low cost of ownership.

The AX400 analyzers are supplied as standard for 85 to 265 V AC operation. There are no inner switches to set.
They can also be provided for 24 V AC or 12 to 30 V DC supply and recognize automatically which of the two supplies is being used. 24 V DC operation reduces maintenance costs significantly by negating the need for costly, yearly safety tests to ensure compliance with safety procedures.

AX460/100010/STDAX460/100010/STD

lf you need to inquire or purchase ,please send the product models to my email or call medirectly .

sunny  He

[Email] sales@xiongbagk.cn

[Mobile] 86-18059884797

[WhatsApp] 86-18059884797

[Skype] sales@saulcontrol.com

AX460/100010/STD Single and dual input analyzers for pH/Redox

Product Description

The backlit display has been designed to operate in all types of environments and shows both the measured parameter(s) and, on a separate 16-character display line, diagnostic and computed information.
On dual-input analyzers both measured parameters are displayed simultaneously.
For conservation of energy, the backlight can be set to switch off automatically after 60 s of inactivity.

The AX400 enables two continuous measurements with simultaneous local display and retransmission. This capability provides an on-line validation of the process measurement. Making two simultaneous measurements and configuring one of the set points as a difference alarm provides continuous on-line checking and early warning of an invalid measurement.

Many installed pH sensors benefit from regular cleaning.
When carried out manually, this can be laborious and costly – so is often ignored. Advanced AX400 pH analyzers make it possible to automate sensor cleaning by using a programmable wash-timer contact to activate a cleaning agent valve or pump.
Pulsed water-wash and jet-wash chemical clean actions are standard, user-selectable functions. The frequency and duration of the wash can be tuned to meet the requirements of the application.
For chemical cleaning a 60 liter (13 gallon) sensor cleaning station can be supplied (model no. AA0010), complete with delivery pump and low level alarm. Chemical cleaning has proved particularly successful on potable drinking water plants where manganese and iron in the sample can create unacceptably high levels of maintenance. Effective cleaning reduces cost of ownership.