Sale!

PXIE-4302 PXI Analog Input Module

A floating signal source is not connected to the building ground system, but has an isolated ground-reference point through the ground of the device. Some examples of floating signal sources are outputs of transformers, thermocouples, battery-powered devices, optical isolators, and isolation amplifiers. An instrument or device that has an isolated output is a floating signal source.
It is important to connect the negative lead of a floating source to AIGND (either directly or through a bias resistor). Otherwise, the source may float out of the maximum working voltage range and return erroneous data. The easiest way to reference the source to AIGND is to connect the positive side of the signal to AI+ and the negative side of the signal to both AIGND and AIwithout using resistors, as shown in Figure 2-1. This connection works well for DC-coupled sources with source impedances of less than 100 Ω

Categories: ,

Description

Overview


Essential details:PXIE-4302 PXI Analog Input Module

A floating signal source is not connected to the building ground system, but has an isolated ground-reference point through the ground of the device. Some examples of floating signal sources are outputs of transformers, thermocouples, battery-powered devices, optical isolators, and isolation amplifiers. An instrument or device that has an isolated output is a floating signal source.
It is important to connect the negative lead of a floating source to AIGND (either directly or through a bias resistor). Otherwise, the source may float out of the maximum working voltage range and return erroneous data. The easiest way to reference the source to AIGND is to connect the positive side of the signal to AI+ and the negative side of the signal to both AIGND and AIwithout using resistors, as shown in Figure 2-1. This connection works well for DC-coupled sources with source impedances of less than 100 Ω

For source impedances ≥100 Ω, this connection leaves the differential signal path significantly off balance. Noise that couples onto the positive line does not couple onto the negative line because it is connected to ground. This noise appears as a differential mode signal instead of a common-mode signal, and therefore appears in your data. In this case, instead of directly connecting the negative line to AIGND, connect the negative line to AIGND through a resistor that is about 100 times the equivalent source impedance, as shown in Figure 2-2. The resistor puts the signal path nearly in balance, so that about the same amount of noise couples onto both connections, yielding better rejection of coupled noise. Since the bias resistor is between the negative line of the floating source and AIGND, this configuration does not load down the floating source output.

PXIE-4302

lf you need to inquire or purchase ,please send the product models to my email or call medirectly .

sunny  He

[Email] sales@saulcontrol.com

[Mobile] 86-18059884797

[WhatsApp] 86-18059884797

[Skype] sales@saulcontrol.com

PXIE-4302 PXI Analog Input Module

Both AI+ and AI- require a DC path to ground. If the source is AC coupled, a resistor is required between the positive input and AIGND, as shown in Figure 2-4. If the source has low-impedance, choose a bias resistor for the positive input that is large enough not to significantly load the source, but small enough not to produce significant input offset voltage as a result of input bias current. Connect the negative input directly to AIGND.

If the source has high output impedance and is AC coupled, balance the signal path as previously described with balanced bias resistors using the same value resistor on both the positive and negative inputs, as shown in Figure 2-5. Be aware that there is some gain error from loading down the source.

A ground-referenced signal source is a signal source connected to a common ground point with respect to the measurement device.
The difference in ground potential between two instruments connected to the same building power system is typically between 1 mV and 100 mV, but the difference can be much higher if power distribution circuits are improperly connected. If a grounded signal source is incorrectly measured, this difference can appear as measurement error. Follow the connection instructions for grounded signal sources to eliminate this ground potential difference from the measured signal.
Figure 2-6 shows how to connect a ground-referenced signal source to an NI PXIe-4302/4303 with a TB-4302.